These ovens feature independent vacuum and purge needle valves.
- 3 in.(7.6cm) glass wool insulation prevents heat loss
- Silicone door gasket and positive latch door maintain seal at all vacuum levels
- Polycarbonate safety shield protects door glass
- Vacuum level is displayed on gauges from 0 to 30 in.Hg
- Vacuum and air lines are corrosion-resistant stainless-steel tubing for optimum chamber cleanliness and long-term performance
- With two removable stacking aluminum shelves
- Easy-to-clean Type 304 stainless-steel interior chamber
- Powder-coated heavy-gauge steel exterior for durability
- Front-mounted three-way valve for evacuation, venting and purging of inert gases (e.g., Nitrogen or Argon) with vacuum fittings on the front of the unit
3606 Models- Hydraulic thermostat temperature control ±1.5°C with a uniformity of ±5°C at 100°C, 25 in.Hg
- Built-in overtemperature protection
- Top-mounted independent evacuation and venting vacuum fittings
3608 Models
- Hydraulic thermostat temperature control ±1.5°C with a uniformity of ±5°C at 100°C, 25 in.Hg
- Built-in overtemperature protection
- Front-mounted three-way valve for evacuation, venting and purging of inert gases (e.g., Nitrogen or Argon) with vacuum fittings on the front of the unit
Model 3618P Models
- Programmable microprocessor-based PID temperature control
- Regulates to ±0.2°C with uniformity of ±2.4°C at 100°C, 25 in.Hg
- Runs at a single setpoint or uses a 24-step program of ramps and dwells
- 0.25 in. compression fittings on left side, require 0.25 in. O.D. copper or stainless-steel tubing
- Front-mounted three-way valve for evacuation, venting and purging of inert gases (e.g., nitrogen or argon) with vacuum fittings on the front of the unit
Ordering Alerts: Digital Units in Torr and Microns available separately (must be factory installed)
Required Accessories: Vacuum fittings require 0.25 in.(0.6cm) I.D. tubing.
Warranty and Service Offering: 12 months