Communication options:
Analog Outputs: 4-20mÅ DC for each stream (optional)
Alarm Outputs: One global dry contact triggered by one or more of the following:
Low sample flow alarm (optional); Low detector flow alarm; Oven/Pyrolyzer temperature fault; Injection valve fault; Purge failure; Calibration fault; Detector temperature fault; Detector lamp voltage fault
One out of service dry contact triggered by: Analyzer in calibration; Suspension of analyzer
Analog Inputs: Optional 4-20mÅ DC inputs from density meter for automatic density compensation of ppm S (w/w); Optional 4-20mÅ DC input from sample flowmeter
Digital Data Communications:
Dual channel with the following optional configurations:
RS-232 Modbus & RS-485 Modbus
Dual channel RS-485 Modbus
TCP/IP encapsulated Modbus & RS-485 Modbus
SOLA II Modbus Remote Interface:
Complete remote control of SOLA; automatic logging of analysis results and analyzer parameters;
communication to SOLA II via serial or TCP/IP encapsulated Modbus enables remote diagnostics via modem
SOLAWeb Remote Interface:
Complete remote control of SOLA; ability to download 24 hours of analysis results and analyzer parameters;
communication to SOLA II via local area network (TCP/IP) enables remote diagnostics via modem
Local MMI (Man Machine Interface):
Status of all analyzer parameters (e.g., furnace & oven temperatures, PMT and lamp voltage, detector flow rate, etc.) and analytical results available on front mounted displays, push button menu access, hazardous area classification remains intact while operating local display
Certifications: (Built to) NEC Class I, Division 2, Groups B, C & D
(Built to) NEC Class I, Division 1 (optional), Groups B, C & D
CSA with associated “C/US Mark“ Class I, Division 2, Groups B, C & D
CSA with associated “C/US Mark“ Class I, Division 1, Groups B, C & D
ATEX Zone 2, EEx p IIC T2 (T3,T4 Optional)
ATEX Zone 1 (optional), EEx p IIC T2 (T3,T4 Optional)
CE Mark