Personal Dust Monitor Model PDM3600

Thermo Scientific* Personal Dust Monitor Model PDM3600 is engineered for timely, high-quality personal dust exposure information. It protects the health of miners by ensuring that the shift-average respiratory exposure does not exceed regulatory limits. Providing mine workers and management with the tools to personally monitor and reduce their exposure, the PDM3600 is the first line of defense in preventing long-term health effects.
The PDM3600 provides miners with the capability to assess their own exposure in real time, as opposed to the manual method, which could take 2-3 weeks for test results. This permits miners to take actions to minimize exposure during the work shift. The PDM3600 may also be used as a valuable engineering tool to quickly assess the impact of changes in engineering controls. Using the shortened average period, the effectiveness of these ventilation changes can be quickly evaluated. This is critical to overall dust control as the equipment is repositioned.
 
 
 
 
Part Number Description   Quantity
3600- personal Dust Monitor Model PDM3600
Request Quote
Please Note: Availability may vary by country.  
 
The performance of the PDM3600 has been evaluated by NIOSH in laboratory and underground mine tests. The results indicate that the automated PDM3600 is equal to or better than the current manual dust collection and analysis method used since 1972. The PDM3600 shift-based data is available immediately after the conclusion of a miner's working hours, and shorter term dust exposure data is available continuously.

Key Features:

  • Bright, low power consuming LED primary work light with dual low power secondary LED lights
  • Waterproof cap lamp battery protection
  • Mine-proven power take-off connector
  • PC-based software for data downloading and review of status flags
  • Integral charger with download station of PC interface
Additional Information:

  • The PDM3600 replaces the miner's cap lamp, sample pump and filter holder with a single belt-worn unit.
  • It starts by drawing a continuous sample of air from a miner's breathing zone, it then removes the particles that are larger than respirable in size, and measures the mass of the dust collected on an exchangeable filter.
  • Dust exposure information computed by the monitor is stored internally and is updated every five seconds on the display.
  • Resulting in immediate and accurate end-of-shift dust exposure levels that are equivalent to filter-based models.
  • Battery Operated
The PDM3600 provides three primary, real-time and two secondary, user-initated measurements:

Screen Display Real-time Primary Measurements

  • 30 MIN CONC = Primary current mass concentration (mg/m3) Represents the latest 30-minute running average of dust exposure
  • CUM1 = Primary cumulative mass concentration (mg/m3) Equals the total dust loading since the beginning of the working shift
  • PERCNT OF LIMIT= Indicates percent of exposure limit for primarysample run

Screen Display Secondary User-initiated Measurements

  • 15 MIN CONC = Secondary current concentration (mg/m3) Represents the last 15-minute running average of dust exposure since user-intiated activiation
  • CUM2 = Secondary mass concentration (mg/m3) Represents the dust loading since user initated activiation
Includes:
  • Cap lamp with secondary LED light source and sample inlet
  • Umbilical line containing conductive tubing for sample transport and electrical connections
  • PDM3600 case containing batteries and sample handling system
  • Charging/communication module
  • PC-based software for data management
  • Available, optional high-impact carrying case

Compliance: The PDM3600 is approved for use in the United States only, and meets MSHA intrinsic safety approval for use in underground mines (certified for coal mine dust sample collection) and NIOSH guidelines for Air Sampling and Analytical Method Development and Evaluation.

Certifications: The PDM3600 is for use in the United States and complies with the following requirements:

  • NIOSH approval number TC-74CPDM-01
  • Certified intrinsically safe by the Mine Safety Health Administration (MSHA) (19-A040002-0)
  • EN61326-1:2005
  • FCC Part 15 subpart B
  • UN T1-T8 (shipment of lithium ion batteries)

Recommended for: The PDM3600 was designed under contract with the Centers for Disease Control and Prevention, and the National Institute for Occupational Safety and Health (Pittsburgh Research Laboratory) to arm miners and mine operators with the information to protect personnel from elevated dust concentrations. The development project was the result of a unique partnership involving government, worker representatives and industry, including NIOSH, the Mine Safety and Health Administration, United Mine Workers of America, National Mining Association, Bituminous Coal Operator's Association and Thermo Fisher Scientific.

 
 
 
 
 
Help image

Need Help?


Have a Question?
We're here to help you.

Select Country for Contact Info This website provides global contact information for our products and services. When you select a Contact Country, the contact information that is displayed by default will be for this country.



  • Sales
  • Service
 

In The Spotlight

The Online Library for Enviromental and Process Monitoring Instruments

NewsView more »

AQI Order Status

Check the status of your Air Quality Instruments order.

uspgh-zcom-p2-leg6.fishersci.com:80