false
ThermoScientific
-1
 

EPD Irradiator IRR-1

Thermo Scientific EPD Irradiator IRR-1 is used for checking the constancy of performance of individual Electronic Personal Dosimeters.
Permits verification/calibration check of EPD radiological performance in the field. Gives confidence that an EPD is performing within design specification.
 
 
 
 
Part Number Description   Quantity
EPD1/32000/011 Irradiator for EPD Mk.2 (Irradiates a single EPD)
Request Quote
Please Note: Availability may vary by country.  
 
Uses chlorine-36 and americium-241 sources to deliver beta and gamma radiation for a predetermined time and with excellent repeatability for an individual EPD.

  • Easy-to-operate,compact benchtop unit requires no special shielded facility
  • Compatible with both EPD Mk1 and MK2
  • Allows large numbers of dosemeters to be processed quickly, easily and with no radiation risk to the operator
  • When the IRR-1 is used in conjunction with the Thermo Scientific EPD Maintenance Database software, records can be maintained for the calibration of individual EPDs
Irradiator Performance

  • 15 to 25°C (59 to 77°F) 95% confidence level
  • Am-241 60 keV g: Statistical accuracy better than +2% (irradiation time 2 min.); systematic positional error of better than 2%
  • CI-36 714 keV β (Emax): Statistical and positional accuracy better than +10% (irradiation time)

Electrical and Safety

  • Supply Voltage: 100V +10%, 47 to 63Hz 120V +10%, 47 to 63Hz; 220V +10%, 47 to 63Hz 240V +10%, 47 to 63Hz
  • Power Supply Interruption: Unit will operate normally with a supply interruption of <50ms
  • Fuses: 2 x 250 mA T (time lag), 20 x 5mm (0.8 x 0.2 in.) dia. at 220/240V; 2 x 500 mA T (time lag), 20 x 5mm (0.8 x 0.2 in.) dia. at 100/120V
  • Power Connector: IEC320, (local IEC320 mains cord required outside UK)
  • Safety and EMC Specification: Equipment for indoor use only; IEC 1010-1 Class 1, Installation Category 2, pollution degree 2; BS EN50081-1:1992; BS EN50082-1:1992
Includes: Supplied with a calibration certificate stating the readings obtained with EPDs calibrated with traceability to national standards, and with a reference EPD which is supplied with the irradiator.

 
 
 
 
 