FHT 1700 Iodine Monitors

Thermo Scientific FHT 1700 is designed to monitor gaseous iodine-131 in ambient air or effluent gas continuously.
Boasts excellent selectivity against other iodine isotopes or interfering noble gases.
 
 
 
 
Part Number Description   Quantity
FHT1700 Iodine monitor
Request Quote
Please Note: Availability may vary by country.  
 
Maintenance-free and low-noise gas ring vacuum pump. Operation is user-friendly with measuring electronics operable under Windows*.

  • Application in ambient air monitoring networks or stack emissions monitoring systems
  • Excellent selectivity against other iodine isotopes or interfering noble gases
  • Maintenance-free long term operation only limited by exchange of iodine adsorbent
  • Direct read-out of measured values in physical units of Bq/m3
  • Easy systems integration via RS232 serial interface
  • LAN coupling optionally available
  • Measuring range for iodine concentration 1Bq/h/m3 to 0.5 x 106Bq/h/m3
  • Detector assembly comprising Marinelli beaker with iodine adsorbent and NaI(Tl) scintillation detector
  • Aerosol filter
  • Heated air inlet for optimised iodine retention
  • Measuring electronics FHT8000 operable under Windows
  • Menu-driven calibration procedure with complete track record of efficiency data
  • Air Flow: ~5m3/h

 
 
 
 