Lutetium Test Adapters

Thermo Scientific Lutetium Test Adapters are a smart alternative to conventional check sources and offer a lot of unique advantages to customers using radiation detectors.
Innovative series of test adapters based on high-purity natural Lutetium-Oxide provides users with the ability to “challenge” the functionality of their radiation detectors. Using a chemically pure bulk substance containing the radioisotope in its natural abundance results in a totally constant and homogeneous surface emission rate; unlike other test sources, the Lu Test Adapter possesses activity that is well-defined and reproducible.
Under US DOT or IATA rules for dangerous goods shipments, these adapters are not considered as radioactive material due to their natural origin and low specific activity.
 
 
 
 
Part Number Description   Quantity
4254948 For RadEye G20 and portable NBR-Detectors; 50g
Request Quote
4254949 For FHT 1386 installed industrial raditaion monitors; for 5L and 12L gamma detectors panels; 100g
Request Quote
425068571 For RadEye B20; 9g
Request Quote
425495010 For portal monitors of ASM, SGS series; 200g
Request Quote
425494910 For portal monitors of ASM, SGS series; 100g
Request Quote
425068371 For RadEye AB100; 82g
Request Quote
4254950 For FHT 1386 installed industrial raditaion monitors, for 25L gamma detectors panels; 200g
Request Quote
425067071 For RadEye PRD; 36g
Request Quote
Please Note: Availability may vary by country. Show All
 
  • A 3.7E10 year half-life means no need for error-prone half-life correction and no need for reoccurring purchase of the (decayed) check sources
  • The adapters provide a highly reproducible and uniform activity content of 50 Bq/g (1.3 nCi/g)
  • All test adapters of the same type have virtually the same activity
  • Beta-type adapters provide nearly identical surface emission rates
  • The design of specially shaped enclosures and high-density Lu2O3 ceramics minimizes the required activity for small-size detectors
    • Due to their natural origin and low specific activity, these adapters are not considered as radioactive material in respect to many national regulations
    • Contain the isotope Lu-176 with a 28 times greater half-life than K-40
    • Adapters can be used for performance verification of highly sensitive gamma detectors and pancake-type beta contamination probes
    Test adapters are available in a variety of sizes for use with various radiation measurement instruments:

    Beta/Gamma Test Adapters

    • High-precision, low-energy test adapters for performance verification of the RadEye B20 and other instruments with pancake detector; 9g Lu2O3; D x H: 50 x 3mm (acrylic glass housing); D x H: 40 x 1mm (Lu2O3 ceramics inlet); typical net count rate for RadEye B20: 6 cps
    • Test adapters for large area beta contamination probes with an inherently homogeneous surface emission rate of 0.8 particles per cm2·s - perfect for training and calibration; 80g Lu2O3; 120 x 200 x 5mm total size; 110cm2 (74 x 148mm) (Lu2O3 surface

    Gamma Test Adapter for RadEye PRD

    • Specially shaped lutetium test adapter to match RadEye PRD housing; 36g Lu2O3
    • Typical net count rate for RadEye PRD: 100 cps
    • Indication of “low energy” NBR alarm
    • A carrying case is included

    Universal Gamma Test Adapter for Portable Scintillation Detectors and Gamma Spectroscopy

    • Lutetium test adapter with 50g Lu2O3
    • D x H: 62 x 7mm (aluminium housing); D x H: 55 x 3mm (Lu2O3 ceramics)
    • Typical net count rate for 2 x 2 in. NaI(Tl) detectors: 300 cps
    • Immediate identification of “artifical” NBR alarm for FHZ 672 E-10, net dose rate 30nSv/h [3µR/h]

    Gamma Test Adapters for Radiation Portal Monitors

    • Choose dual disc 100g design or quad-disc 200g design
    • Dual-disc design has a typical net count rate for large area detectors of 1000 cps; 171 x 61 x 17mm (aluminium housing); (2x) D x H: 55 x 3mm (Lu2O3 ceramics)
    • Quad-disc design has a typical net count rate for large area detectors of 2000 cps; recommended for locations with higher background; (4x) D x H: 55 x 3mm (Lu2O3 ceramics)
    • Both styles can be mounted directly to FHT 1388 area detectors (plastic housing) and are also available with holder for monitors with metal enclosure (ASM/SGS installations)

     
     
     
     
     