The Model 15i analyzer operates on the principle that hydrogen chloride (HCl) absorbs infrared radiation at a wavelength of 3.4 microns. Because infrared absorption is a nonlinear measurement technique, it is necessary for the instrument electronics to transform the basic analyzer signal into a linear output. The instrument uses an exact calibration curve to accurately linearize the instrument output over any range up to a concentration of 5000ppm in emissions from a source.
Key Features:- Gas Filter Correlation Technology
- Self-aligning optics
- Standard dual range and auto ranges
- Temperature and pressure correction
- User selectable alarm levels for concentration
iSeries Advantage: - Ethernet connectivity for efficient remote access
- Enhanced user interface with one button programming and large display screen
- Flash memory for increased data storage and user downloadable software
- Enhanced electronics design optimizes product commonality
- Improved layout for easier accessibility to components
Ordering Alerts: Voltage optionsA: 120VAC 50/60Hz (standard)
B: 220VAC 50/60Hz
J: 100VAC 50/60Hz
Internal zero /span
N: No zero / span valve
Z: Internal zero / span valves
Filter Wheel Purge
S: Standard plumbing (standard)
P: Filter wheel purge set-up
Optional I/O
A: None (standard)
C: I/O expansion board (4 to 20mA outputs - 6 channels, 0 to 10V inputs - 8 channels)
Mounting Hardware
A: Bench mounting (standard)
B: Ears and handles, EIA
C: Ears and handles, Retrofit
Compliance: ETL, CSA, UL, CE