| Specifications for Model 43i SO2 Analyzer |
| Measurement Ranges | Preset Ranges 0-0.05, 0.1, 0.2, 0.5, 1, 2, 5, and 10ppm 0-0.2, 0.5, 1, 2, 5, 10, 20, and 25mg/m3 Extended Ranges 0-0.05, 1, 2, 5, 10, 20, 50 and 100ppm 0-2, 5, 10, 20, 50, 100, 200, and 250mg/m3 Custom Ranges 0-0.05 to 100ppm 0-0.2 to 250mg/m3 |
| Interferences | Interferences (EPA levels) : < lower detectable limit except for the following: NO < 3ppb, M-Xylene < 2ppb, H2O < 2% of reading |
| Flowrate | 0.5L/min. (standard); 1L/min. (optional) |
| Min. Detection Limit | 2.0ppb (10 second averaging time), 1.0ppb (60 second averaging time), 0.5ppb (300 second averaging time) |
| Drift | Zero Drift (24 hour): less than 1ppm per day Span Drift (24 hour): ±1% |
| Linearity | ±1% full scale < 100ppm |
| Precision | 1% of reading or 1ppb (whichever is greater) |
| Zero Noise | 1.0ppb RMS (10 second averaging time), 0.5ppb RMS (60 second averaging time), 0.25ppb RMS (300 second averaging time) |
| Response Time | 80 seconds (10 second average time) 110 seconds (60 second average time) 320 seconds (300 second average time) |
| Power Requirements | 100VAC, 115VAC, 220-240VAC ±10%, 165w |
| Input/Output | Outputs : Selectable Voltage, RS232/RS485, TCP/IP, 10 Status Relays, and Power Fail Indication (standard). 0 to 20 or 4 to 20mA Isolated Current Outout (optional) Inputs : 16 Digital Inputs (standard), eight 0-10VDC Analog Inputs (optional) |
| Temperature Range [Operating] | 20° to 30°C (68° to 86°F) |
| Dimensions | D x W x H: 23 x 16.75 x 8.62 in. (584 x 425 x 219mm) |
| Weight | 48 lb. (21.8kg) |